Design and use of computational models for contrast-enhanced spectral mammography

Yanka Baneva, Tihomir Georgiev, Kristina Bliznakova

Abstract

Introduction: Contrast-enhanced spectral mammography (CESM) is a new technique for cancer investigations. The application of this technique to screen the breast would result in contrast improvement of the different breast lesions. This can be effectively ensured by the use of phantoms. Computational phantoms are one of the ways to investigate the characteristics of the received image and thus to evaluate the whole technique.

Aim: The aim of this study is to validate and analyse the design of three different computational phantoms for CESM.

Materials and Methods: An in-house software tool was used to create three computational phantoms, consisting of iodinated inserts for the simulation of the CESM procedure. The inserts in the phantoms, modelled from Omnipaque, have a radius of 8–10 mm and varying height. The three phantoms are made of polymethyl methacrylate (PMMA), with different shape and composition. One of the phantoms is characterised by a heterogeneous background. For each phantom two x-ray radiographs were generated, one at x-ray energy of 20keV and one at 34 keV. The images were processed to obtain a recombined iodine image, which shows the iodine contrast agent and suppresses the surrounding background tissue.

Results: Simulated spectral images demonstrated a great improvement of the image quality compared to low-energy images of the phantoms. The simulations with the inhomogeneous model revealed that the heterogeneous background has been successively depressed while improving the visibility of the iodine inserts.

Conclusion: The heterogeneous breast phantom might be used as a reference tool for information about the needed iodine concentration which needs to be inserted during the procedure to obtain significant enhancement in the suspicious area.


Keywords

computational breast models, contrast-enhanced spectral mammography, iodine concentration, modelling, simulation

Full Text


References

Heck L, Dierolf M, Jud C, Eggl E, Thorsten S, Mechlem K, et al. Contrast-enhanced spectral mammography with a compact synchrotron source. Roeder RK, ed. PLoS ONE. 2019;14(10):e0222816. doi:10.1371/journal.pone.0222816.

Carton AK, Saab-Puong S, Suminski M. SenoBright Contrast Enhanced Spectral Mammography Technology. Published online 2012.

Mohamed Kamal R, Hussien Helal M, Wessam R, Mahmoud Mansour S, Godda I, Alieldin N. Contrast-enhanced spectral mammography: Impact of the qualitative morphology descriptors on the diagnosis of breast lesions. Eur J Radiology. 2015;84(6):1049-5. doi:10.1016/j.ejrad.2015.03.005.

Luczynska E, Niemiec J, Heinze S, Adamszyk A, Ambicka A, Marcyniuk P, et al. Intensity and pattern of enhancement on CESM: prognostic significance and its relation to expression of podoplanin in tumor stroma – a preliminary report. Anticancer Res. 2018;38(2):1085-95. doi:10.21873/anticanres.12327.

Mammo CESMTM Phantom (GammexTM Technology) - Sun Nuclear. Available from: https://www.sunnuclear.com/products/mammo-cesm-phantom

CIRS Releases Contrast Enhanced Spectral Mammography Phantom. CIRS. Available from: https://www.cirsinc.com/news/cirs-releases-contrast-enhanced-spectral-mammography-phantom-a-simple-comprehensive-phantom-for-routine-cesm-qa/

Baldelli P, Bravin A, Di Maggio C, Gennaro G, Gambaccini M, Sarnelli A, et al. Evaluation of the minimum iodine concentration for contrast enhanced subtraction mammography. Nucl Instrum Methods Phys Res A. 2007;580(2):1115-8. doi:10.1016/j.nima.2007.06.109.

Contillo A, Taibi A. [P053] Contrast-enhanced digital mammography on commercial systems: an optimization study. Phys Med. 2018;52:114. doi:10.1016/j.ejmp.2018.06.376.

Bliznakova K, Speller R, Horrocks J, Liaparinos P, Kolitsi Z, Pallikarakis N. Experimental validation of a radiographic simulation code using breast phantom for X-ray imaging. Computers in Biology and Medicine. 2010;40(2):208-14. doi:10.1016/j.compbiomed.2009.11.017.

Bliznakova K, Buliev I, Bliznakov Z. Anthropomorphic Phantoms in Image Quality and Patient Dose Optimization: A EUTEMPE Network Book. IOP Publishing; 2018.

Baneva Y, Bliznakova K, Cockmartin L, Marinov S, Buliev I, Mettivier G, et al. Evaluation of a breast software model for 2D and 3D X-ray imaging studies of the breast. Phys Med. 2017;41:78-86. doi:10.1016/j.ejmp.2017.04.024.

Cockmartin L, Bosmans H, Bliznakova K, Pokrajac D, Imran A, Marshall N, et al. Creation of Realistic Structured Backgrounds using Adipose Compartment Models in a Test Object for Breast Imaging Performance Analysis. Published online 2016.

Rudnicki W, Heinze S, Niemiec J, Kojs Z, Sas-Korczynska B, Hendrick E, et al. Correlation between quantitative assessment of contrast enhancement in contrast-enhanced spectral mammography (CESM) and histopathology—preliminary results. Eur Radiol. 2019;29(11):6220-6226. doi:10.1007/s00330-019-06232-6.

Bliznakova K, Dukov N, Feradov F, Gospodainova G, Bliznakov Z, Russo P, et al. Development of breast lesions models database. Physica Medica. 2019;64:293-303. doi:10.1016/j.ejmp.2019.07.017.




DOI: http://dx.doi.org/10.14748/ssm.v54i2.8361
Array
Article Tools
Email this article (Login required)
About The Authors

Yanka Baneva
Medical University of Varna
Bulgaria

Department of Physics and Biophysics, Faculty of Pharmacy

Tihomir Georgiev
Medical University of Varna
Bulgaria

Department of Medical Equipment, Electronic and Information Technologies in Healthcare, Faculty of Public Health

Kristina Bliznakova
Medical University of Varna
Bulgaria

Department of Medical Equipment, Electronic and Information Technologies in Healthcare, Faculty of Public Health

Font Size


|